1.National Key Laboratory of Laser Spatial Information, Harbin Institute of Technology, Harbin 150001, China
2.Zhengzhou Research Institute, Harbin Institute of Technology, Zhengzhou 450000, China
Yufei Ma (mayufei@hit.edu.cn)
收稿日期:2025-02-04,
修回日期:2025-04-02,
录用日期:2025-04-08,
网络出版日期:2025-04-30,
纸质出版日期:2025-07-31
Scan QR Code
Parts-per-quadrillion level gas molecule detection: CO-LITES sensing[J]. LSA, 2025,14(7):1882-1892.
Sun, H. Y. et al. Parts-per-quadrillion level gas molecule detection: CO-LITES sensing. Light: Science & Applications, 14, 1882-1892 (2025).
Parts-per-quadrillion level gas molecule detection: CO-LITES sensing[J]. LSA, 2025,14(7):1882-1892. DOI: 10.1038/s41377-025-01864-4.
Sun, H. Y. et al. Parts-per-quadrillion level gas molecule detection: CO-LITES sensing. Light: Science & Applications, 14, 1882-1892 (2025). DOI: 10.1038/s41377-025-01864-4.
Highly sensitive gas detection plays a crucial role in advanced scientific and technological fields. This paper presents a parts-per-quadrillion (ppq) level ultra-highly sensitive light-induced thermoelectric spectroscopy (LITES) sensor for the first time. The artificial fish swarm algorithm auto-designed multi-pass cell (MPC) with double helix pattern
and the polymer modified round-head quartz tuning fork (QTF) with low-resonant frequency (
f
0
) were adopted to improve the gas absorption and QTF's detection ability. The obtained MPC
with a long optical path length (OPL) of 25.8 m and a small volume of 165.8 ml
is beneficial for increasing gas absorption while keeping the sensor compact. The novel QTF was structurally optimized to obtain low
f
0
(~9.5 kHz) and modified by polydimethylsiloxane (PDMS) to reduce heat diffusion and enhance vibration amplitude. A strong absorption line of carbon monoxide (CO) located in the mid-infrared region (4.59 μm) was chosen as the target line. The signal-to-noise ratio (SNR) of CO-LITES sensor based on the novel QTF was improved by 10.59 times
reaching the highest level when compared to the commercial QTF. The corresponding minimum detection limit (MDL) was calculated to be 23 ppt. When the integration time of the sensor system was increased to 500 s
the MDL could be improved to 920.7 ppq. Compared to the reported spectroscopy techniques for CO gas detection
the LITES sensor in this study offers an excellent result in terms of detection sensitivity.
Yu, Q. Z. et al. Transient stimulated Raman scattering spectroscopy and imaging. Light Sci. Appl. 13 , 70 (2024)..
Wang, X. Y. et al. Flat soliton microcomb source. Opto-Electron. Sci. 2 , 230024 (2023)..
Dong, L. et al. Compact CH 4 sensor system based on a continuous-wave, low power consumption, ro om temperature interband cascade laser. Appl. Phys. Lett. 108 , 011106 (2016)..
Fan, J. J., Ou, Z. Y. & Zhang, Z. D. Entangled photons enabled ultrafast stimulated Raman spectroscopy for molecular dynamics. Light Sci. Appl. 13 , 163 (2024)..
Yi, H. M. et al. Short-lived species detection of nitrous acid by external-cavity quantum cascade laser based quartz-enhanced photoacoustic absorption spectroscopy. Appl. Phys. Lett. 106 , 101109 (2015)..
Dong, S. L. et al. Early cancer detection by serum biomolecular fingerprinting spectroscopy with machine learning. eLight 3 , 17 (2023)..
Fjodorow, P. et al. Room-temperature Fe: ZnSe laser tunable in the spectral range of 3.7-5.3 µm applied for intracavity absorption spectroscopy of CO 2 isotopes, CO and N 2 O. Opt. Express 29 , 12033–12048 (2021)..
Minamikawa, T. et al. Long-range enhancement for fluorescence and Raman spectroscopy using Ag nanoislands protected with column-structured silica overlayer. Light Sci. Appl. 13 , 299 (2024)..
Zhang, M. H. et al. Theoreti cal design of MOFs and PSAprocess for efficient separation of CF 4 /NF 3 . Ind. Eng. Chem. Res. 62 , 7103–7113 (2023)..
Cao, L. N. et al. Atomically dispersed iron hydroxide anchored on Pt for preferential oxidation of CO in H 2 . Nature 565 , 631–635 (2019)..
Korablev, O. et al. No detection of methane on Mars from early ExoMars trace gas orbiter observations. Nature 568 , 517–520 (2019)..
Shao, M. R. et al. Ferroelectrically modulate the Fermi level of graphene oxide to enhance SERS response. Opto-Electron. Adv. 6 , 230094 (2023)..
Qiao, S. D. et al. Ultra-highly sensitive dual gases detection based on photoacoustic spectroscopy by exploiting a long-wave, high-power, wide-tunable, single-longitudinal-mode solid-state laser. Light Sci. Appl. 13 , 100 (2024)..
Chen, W. P. et al. Mid-infrared all-fiber light-induced thermoelastic spectroscopy sensor based on hollow-core anti-resonant fiber. Photoacoustics 36 , 100594 (2024)..
Ma, Y. F. et al. A novel tapered quartz tuning fork-based lase r spectroscopy sensing. Appl. Phys. Rev. 11 , 041412 (2024)..
Li, C. G. et al. Compact TDLAS based optical sensor for ppb-level ethane detection by use of a 3.34 µm room-temperature CW interband cascade laser. Sens. Actuators B: Chem. 232 , 188–194 (2016)..
Yang, Q. F. et al. Efficient microresonator frequency combs. eLight 4 , 18 (2024)..
Wang, R. Q. et al. Highly sensitive laser spectroscopy sensing based on a novel four-prong quartz tuning fork. Opto-Electron. Adv. 8 , 240275 (2025)..
Lang, Z. T. et al. Disturbance-immune, fast response LITES gas sensor based on out-plane vibration mode employing micro Fabry-Perot cavity with heterodyne phase demodulation. Sens. Actuators B: Chem. 419 , 136412 (2024)..
Hashimoto, K. et al. Upconversion time-stretch infrared spectroscopy. Light Sci. Appl. 12 , 48 (2023)..
Kosterev, A. A. et al. Quartz-enhanced photoacoustic spectroscopy. Opt. Lett. 27 , 1902–1904 (2002)..
Ma, Y. F. et al. HCl ppb-level detection based on QEPAS sensor using a low resonance frequency quartz tuning fork. Sens. Actuators B: Chem. 233 , 388–393 (2016)..
Wu, H. P. et al. Quartz enhanced photoacoustic H 2 S gas sensor based on a fiber-amplifier source and a custom tuning fork with large prong spacing. Appl. Phys. Lett. 107 , 111104 (2015)..
Ma, Y. F. et al. Quartz-tuning-fork enhanced photothermal spectroscopy for ultra-high sensitive trace gas detection. Opt. Express 26 , 32103–32110 (2018)..
Zhang, C. et al. High-sensitivity trace gas detection based on differential Helmholtz photoacoustic cell with dense spot pattern. Photoacoustics 38 , 100634 (2024)..
Hu, L. E. et al. Compact all-fiber light-induced thermoelastic spectroscopy for gas sensing. Opt. Lett. 45 , 1894–1897 (2020)..
Ma, Y. F. et al. Highly sensitive and fast hydrogen detection based on light-induced thermoelastic spectroscopy. Ultrafast Sci. 3 , 0024 (2023)..
Zhang, D. Q. et al. Cavity-enhanced light-induced thermoelastic spectroscopy for trace-gas sensing. Opt. Express 32 , 33618–33627(2024)..
Hu, L. E. et al. Long-distance in-situ methane detection using near-infrared light-induced thermo-elastic spectroscopy. Photoacoustics 21 , 100230 (2021)..
Liu, Y. H. & Ma, Y. F. Advances in multipass cell for absorption spectroscopy-based trace gas sensing technology. Chin. Opt. Lett. 21 , 033001 (2023)..
Dong, M. et al. Double-range near-infrared acetylene detection using a dual spot-ring Herriott cell (DSR-HC). Opt. Express 26 , 12081–12091 (2018)..
He, Y. et al. Hydrogen-enhanced light-induced thermoelastic spectroscopy sensing. Photonics Res. 13 , 194–200 (2025)..
Sun, B. et al. Light-induced thermoelastic sensor for ppb-level H 2 S detection in a SF 6 gas matrices exploiting a mini-multi-pass cell and quartz tuning fork photodetector. Photoacoustics 33 , 100553 (2023)..
Li, C. L. et al. Optical heterodyne Herriott-type multipass laser absorption spectrometer . Chin. Opt. Lett. 13 , 013001 (2015)..
Luo, H. J. et al. Ppbv-level mid-infrared photoacoustic sensor for mouth alcohol test after consuming lychee fruits. Photoacoustics 33 , 100559 (2023)..
Menduni, G. et al. High-concentration methane and ethane QEPAS detection employing partial least squares regression to filter out energy relaxation dependence on gas matrix composition. Photoacoustics 26 , 100349 (2022)..
Li, B. et al. Calibration-free mid-infrared exhaled breath sensor based on BF-QEPAS for real-time ammonia measurements at ppb level. Sens. Actuators B: Chem. 358 , 131510 (2022)..
Pinto, D. et al. Parts-per-billion detection of carbon monoxide: a comparison between quartz-enhanced photoacoustic and photothermal spectroscopy. Photoacoustics 22 , 100244 (2021)..
Liu, X. N. & Ma, Y. F. Sensitive carbon monoxide detection based on light-induced thermoelastic spectroscopy with a fiber-coupled multipass cell [invited ] . Chin. Opt. Lett. 20 , 031201 (2022)..
Patimisco, P. et al. A quartz enhanced photo-acoustic gas sensor based on a custom tuning fork and a terahertz quantum cascade laser. Analyst 139 , 2079–2087 (2014)..
Qiao, S. D. et al. Ultra-highly sensitive HCl-LITES sensor based on a low-frequency quartz tuning fork and a fiber-coupled multi-pass cell. Photoacoustics 27 , 100381 (2022)..
Russo, S. D. et al. Quartz-enhanced photoacoustic spectroscopy exploiting low-frequency tuning forks as a tool to measure the vibrational relaxation rate in gas species. Photoacoustics 21 , 100227 (2021)..
Li, S. Z. et al. Ppb-level quartz-enhanced photoacoustic detection of carbon monoxide exploiting a surface grooved tuning fork. Anal. Chem. 91 , 5834–5840 (2019)..
Kwon, B. et al. Dynamic thermomechanical response of bimaterial microcantilevers to periodic heating by infrared radiation. Rev. Sci. Instrum. 83 , 015003 (2012)..
Pushkarsky, M. B. et al. High-sensitivity detection of TNT. Proc. Natl Acad. Sci. USA 103 , 19630–19634 (2006)..
Werle, P. O., Mücke, R. & Slemr, F. The limits of signal averaging in atmospheric trace-ga s monitoring by tunable diode-laser absorption spectroscopy (TDLAS). Appl. Phys. B 57 , 131–139 (1993)..
Cui, R. Y. et al. Generalized optical design of two spherical-mirror multi-pass cells with dense multi-circle spot patterns. Appl. Phys. Lett. 116 , 091103 (2020)..
Kong, R. et al. Optical design and analysis of a two-spherical-mirror-based multipass cell. Appl. Opt. 59 , 1545–1552 (2020)..
Tuzson, B. et al. Compact multipass optical cell for laser spectroscopy. Opt. Lett. 38 , 257–259 (2013)..
Graf, M., Emmenegger, L. & Tuzson, B. Compact, circular, and optically stable multipass cell for mobile laser absorption spectroscopy. Opt. Lett. 43 , 2434–2437 (2018)..
Yin, X. K. et al. Near-infrared laser photoacoustic gas sensor for simultaneous detection of CO and H 2 S. Opt. Express 29 , 34258–34268 (2021)..
Sgobba, F. et al. Compact and portable quartz-enhanced photoacoustic spectroscopy sensor for carbon monoxide environmental monitoring in urban areas. Photoacoustics 25 , 1 00318 (2022)..
Shao, L. G. et al. Highly precise measurement of atmospheric N 2 O and CO using improved white cell and RF current perturbation. Sens. Actuators B: Chem. 352 , 130995 (2022)..
Kloss, C. et al. Airborne mid-infrared cavity enhanced absorption spectrometer (AMICA). Atmos. Meas. Tech. 14 , 5271–5297 (2021)..
0
浏览量
0
Downloads
0
CSCD
关联资源
相关文章
相关作者
相关机构
京公网安备11010802024621